Monitoring vacuum processes with a conventional RGA at pressures >10-4 Torr typically requires the addition of differential pumping. Alternative analyser types optimised for high pressure operation have degraded sensitivity at low pressures.

The innovative Hiden HMT analyser enables operation at high pressure yet maintains full RGA performance at high vacuum with dual mode operation:

  • UHV mode for high performance residual gas analysis at pressure <10-4 Torr through to 10-13 Torr
  • high pressure mode for measurement at pressures >10-4 Torr through to 5X10-3 Torr

The perfomance of the HMT system operating in high pressure mode compared to regular RGA performance is illustrated in the graphs below where Argon fill gas flow is increased progressively:

HV0002_Figure1_450px

Figure 1. HMT analyser – measuring Argon in the chamber up

HV0002_Figure2_450px

Figure 2. Chamber pressure

HV0002_Figure3_450px

Figure 3. The same experiment monitored by a regular RGA. Non linearity occurs for measurements at pressure > 3X10-5 Torr

Visit Product Page : HMT

View Performance Data Sheet :  High Pressure RGA Process monitoring with the Hiden HMT dual mode high pressure RGA